Phenomenon |
Cause |
Solution |
1. No power indication |
1.The power cord is not plugged in or disconnected 2.Open fuse |
1.Insert the power plug, check the power cord 2. Replace the fuse |
2. The temperature in the box does not rise |
1. The set temperature is too low 2. The temperature controller is damaged. 3.The electric heater is damaged |
1. Adjust the set temperature 2. Replace the temperature controller 3. Replace the electric heater
|
3. The PV of the meter displays "0000" or the numbers are string up and down. |
The temperature sensor is damaged. |
Replace the temperature sensor. |
4. Abnormal over temperature alarm |
1. The set temperature is too low 2. The temperature controller is damaged |
1. Adjust the set temperature 2.Replace the temperature controller
|
5. Humidity does not go up |
Over-temperature setting is too low |
Set over-temperature value correctly, between 55-60 |
The fan does not rotate |
Check and confirm that the fan resistance is too large or the capacitor is damaged, and replace the fan or capacitor of the same type |
|
There is a problem with the sensor |
Replace the sensor of the same type |
|
The water level of the water tank is not enough; the position of the water cup is high. |
Fill up with water; Adjust the position of the water cup
|
|
There is no water in the humidification shallow water tank, and the water level alarm will act. |
Check the waterway to make the waterway unblocked. The float action should have on-off action, without replacing the float. |
|
Wet bulb sensor is deviated |
Replace sensor |
|
6. The evaporator is frosted or the inner liner is frosted. |
Open the door frequently when it is at low temperature. |
Stop cooling and dry it at 80 degrees, which may reduce the number of door openings. |
The lead hole is not tightly sealed |
Cover with rubber stopper during low temperature
|